PROCEDURE: 3h TITLE: Internal calibration lamp flats (HRC) CATEGORY: Verification PURPOSE: (1) Test whether tungsten calibration lamp works and provides uniform illumination. (2) Test spectral model for Tungsten lamps. Only need to do this for a few broad band filters FREQUENCY: Run once for each lamp at nominal environmental T. DETECTOR: HRC PREREQUISITES: none HARDWARE REQUIREMENTS: none SOFTWARE REQUIREMENTS: none COMMAND MODE: SMS (done) PROCEDURE/EXPOSURES: GAIN = 2 Turn tungsten lamp on (see variants below) 1. bias 0s 2. F250W 600s (probably will get only ~1100 e-) 3. SDSS g 5s 4. SDSS r 1s 5. SDSS i 0.5s 6. SDSS z 0.5s 7. bias 0s VARIANTS: a- Tungsten #3 JTVH3A_1.SMS b- Tungsten #4 JTVH3B_1.SMS TOTAL EXPOSURES: 7 TOTAL TIME: setup + 7*overhead + 607s = 0.57hr (setup = 10min, overhead = 2min) ANALYSIS: standard statistical analysis. Compare count rate to model.