PROCEDURE: 3w TITLE: Internal calibration lamp flats (WFC) CATEGORY: Verification PURPOSE: (1) Test whether tungsten calibration lamp works and provides uniform illumination. (2) Test spectral model for Tungsten lamps. Only need to do this for a few broad band filters FREQUENCY: Run once for each lamp at nominal environmental T. DETECTOR: WFC PREREQUISITES: none HARDWARE REQUIREMENTS: none SOFTWARE REQUIREMENTS: none COMMAND MODE: SMS PROCEDURE/EXPOSURES: GAIN = 2 Turn tungsten lamp on (see variants below) 1. bias 0s 2. SDSS g 2.0s 3. SDSS r 0.5s 4. SDSS i 0.7s 5. SDSS z 0.7s 6. bias 0s VARIANTS: a- Tungsten #1 b- Tungsten #2 TOTAL EXPOSURES: 6 TOTAL TIME: setup + 6*overhead + 3.9s = 1.17 hr (setup = 10min, overhead = 10min) ANALYSIS: standard statistical analysis. Compare count rate to model.