PROCEDURE: 2h TITLE: CTE monitor (HRC) CATEGORY: monitor PURPOSE: Check charge transfer properties of CCDs FREQUENCY: Run after every major change, e.g. temp, at 5 exp. levels. Run this set once with serial and once with parrallel read out. DETECTOR: HRC PREREQUISITES: (1) Verify count rate of tungsten#3 lamp with the F502N and F625W filters. Do this preferabley outside of TV. - Done. RAS/HOMS testing (11/98) shows count rates with Tungsten 3 lamp, 4.0 times larger than model. Exposure times used below are made using revised count-rates of C = 206, 2.56e4 e/s for F502N, F625W. (2) Verify FPR read-out method with HRC using flight software. HARDWARE REQUIREMENTS: none SOFTWARE REQUIREMENTS: FPR readout software M.R. Jones IDL FPR software. COMMAND MODE: SMS PROCEDURE/EXPOSURES: GAIN=1 Tungsten lamp (#3) on 1. bias 0.0s -- read exposures 2-6 with serial FPR timing 2. F502N 0.5s (100) 3. F502N 2.5s (500) 4. F502N 5.0s (1000) 5. F625W 0.2s (5000) 6. F625W 0.4s (10000) -- read exposures 7-11 with serial FPR timing 7. F502N 0.5s (100) 8. F502N 2.5s (500) 9. F502N 5.0s (1000) 10. F625W 0.2s (5000) 11. F625W 0.4s (10000) 12. bias 0.0s Tungsten lamps #3 off VARIANTS: a- both serial & parrallel FPR sets b- Serial FPR readout set only c- Parrallel FPR readout readout set only TOTAL EXPOSURES: 12 (variant a), 7 (variants b,c) TOTAL TIME: setup + (7 or 12)*overhead + (5 or 10)*(<~10s) = 0.59: hr (variant a) = 0.41: hr (variants b,c) (setup = 10min, overhead = 2min) ANALYSIS: (1) Determine CTE at different levels. Use M.R. Jones' FPR software. (2) Compare with results at other T_CCD, thorough CTE test.