PROCEDURE: 2w TITLE: CTE monitor (WFC) CATEGORY: monitor PURPOSE: Check charge transfer properties of CCDs FREQUENCY: Run after every major change, e.g. temp, gain setting, at 5 exp. levels. Run this set once with parallel read-out and EPER software, and once with serial read-out and FPR software. DETECTOR: WFC PREREQUISITES: none HARDWARE REQUIREMENTS: none SOFTWARE REQUIREMENTS: EPER & FPR readout software M.R. Jones IDL EPER and FPR software. COMMAND MODE: SMS? PROCEDURE/EXPOSURES: GAIN = 1 Tungsten lamp (#1) on 1. bias 0.0s -- read exposures 2-6 with serial FPR timing 2. F555W x F435W 8.0s (500) 3. F555W x F435W 16.0s (1000) 4. F435W 1.2s (2000) 5. F435W 3.0s (5000) 6. F435W 12.0s (20000) -- read exposures 7-11 with parallel EPER timing 7. F555W x F435W 8.0s (500) 8. F555W x F435W 16.0s (1000) 9. F435W 1.2s (2000) 10. F435W 3.0s (5000) 11. F435W 12.0s (20000) -- read last exposure with normal read out 12. bias 0.0s VARIANTS: a- both serial FPR & parallel EPER readouts b- serial FPR readouts only c- parrallel EPER readout TOTAL EXPOSURES: a- 12 b,c- 7 TOTAL TIME: setup + (7 or 12)*overhead + (40.2s or 80.4s) = 2.19 variant a = 1.34 variants b,c (setup = 10min, overhead = 10min) ANALYSIS: (1) Determine CTE at different levels using EPER and FPR CTE analysis software. (2) Compare with results at other T_CCD, thorough CTE test.