PROCEDURE: 6h TITLE: Long darks CATEGORY: Calibration PURPOSE: Measure darks with long enough exposures to get significant counts. FREQUENCY: Run over range of nominal operating T. Use as filler when not doing anything else. DETECTOR: HRC PREREQUISITES: none HARDWARE REQUIREMENTS: none SOFTWARE REQUIREMENTS: standard statistical analysis tool stsdas.hst_calib.ctools.msbadpix (or IDL equiv) COMMAND MODE: PSTOL PROCEDURE/EXPOSURES: 1. bias 0s 2. dark 3600s (repeat as needed) Note T_env, T_CCD at beginning, end and in the middle of each dark exposure. TOTAL EXPOSURES: 2 TOTAL TIME: setup + 3600s + 2*overhead = 1.23 hr (+ 1.03hr for each additional dark) (setup = 10min, overhead = 2min) ANALYSIS: (1) Subtract overscan or master bias. (2) Run basic statistical analysis tool. Gives dark count rate. Compare with previous darks to check if rate & shape remain constant. (3) Plot dark rate vs. T_env, T_CCD, time. Note trends. (4) If more than one dark, combine and identify bad pixels.