Table 2a (old). Throughput measurements of Kr 1236Å (no bad row replacement)
Entry # MJD -
51200
Ilamp
[Amp]
SBC
filter
texp
[s]
global rate
[counts/s]
SBC/PMT
(rate ratio)
notes
Splat dark
3472 3473 29.09001 0.15 F115LP 100 1828 ± 4 0.719 ± 0.003 1  
3474 3473 29.10709 0.15 F115LP 100 1822 ± 4 0.695 ± 0.003 1  
3620 3619 31.85814 0.15 F115LP 100 1821 ± 4 0.648 ± 0.003 1  
3621 3622 31.86462 0.15 F115LP 100 1796 ± 4 0.642 ± 0.003 1  
3694 3695 34.69314 0.15 F115LP 100 2218 ± 4 0.702 ± 0.002  
3696 3695 34.70228 0.15 F115LP 100 2220 ± 4 0.709 ± 0.002  
3828 3829 37.00398 0.15 F115LP 100 2215 ± 4 0.694 ± 0.002 2a  
3830 3829 37.01102 0.15 F155LP 100 2203 ± 4 0.682 ± 0.002 2a  
4304 4305 39.06843 0.15 F115LP 100 2185 ± 4 0.720 ± 0.002  
4306 4305 39.07499 0.15 F115LP 100 2182 ± 4 0.725 ± 0.002  
4799 4800 42.14994 0.15 F115LP 100 2132 ± 4 0.694 ± 0.002  
4801 4800 42.15883 0.15 F115LP 100 2167 ± 4 0.700 ± 0.002  
4857 4858 44.40812 0.15 F115LP 100 2170 ± 4 0.712 ± 0.002  
4859 4858 44.41620 0.15 F115LP 100 2165 ± 4 0.714 ± 0.002  
4925 4926 47.11604 0.20 F115LP 300 2479 ± 3 0.583 ± 0.001 1  
4927 4926 47.12706 0.20 F125LP 500 601 ± 1 0.143 ± 0.003 1  
4928 4929 47.13641 0.20 F122M 500 392 ± 1 0.094 ± 0.004 1  
4930 4929 47.14935 0.20 PR130L 500 603 ± 1 0.147 ± 0.003 1  
4931 4932 47.15863 0.20 PR110L 500 1272 ± 1 0.312 ± 0.002 1  
4933 4932 47.17149 0.15 F115LP 500 2018 ± 2 0.707 ± 0.002 1  
4934 4932 47.18228 0.15 F125LP 500 503 ± 1 0.176 ± 0.003 1  
4935 4932 47.19093 0.15 F122M 500 327 ± 1 0.114 ± 0.004 1  
4936 4938 47.19980 0.15 PR130L 500 514 ± 1 0.179 ± 0.003 1  
4937 4938 47.20813 0.15 PR110L 500 1079 ± 1 0.376 ± 0.002 1  
5897 5898 51.38228 0.15 F115LP 300 1962 ± 2 0.628 ± 0.002 1 2b  
5899 5900 51.39399 0.15 F115LP 100 1960 ± 4 0.633 ± 0.003 2b  
5901 5900 51.40519 0.20 F115LP 100 2355 ± 5 0.487 ± 0.002 2b  
5902 5900 51.41447 0.10 F115LP 100 3078 ± 5 0.822 ± 0.002 2b  
5903 5904 51.42307 0.15 F115LP 100 19283 ±13 0.618 ± 0.001 1 2b  
5905 5904 51.43681 0.15 F115LP 100 18343 ±13 0.624 ± 0.001 1 2b  
5908 5907 51.46587 0.15 F115LP 100 8428 ± 9 0.627 ± 0.001 2b 3  
5909 5907 51.47022 0.15 F115LP 100 1977 ± 4 0.649 ± 0.003 2b  
5910 5911 51.47654 0.15 F115LP 100 14214 ±12 0.638 ± 0.001 1 2b 4  
5912 5911 51.49079 0.05 F115LP 100 6162 ± 8 0.833 ± 0.001 1 2b  
5913 5914 51.50111 0.08 F115LP 100 1609 ± 4 0.829 ± 0.003 1 2b  
Notes:
1.  Dark radiometer reading not properly obtained. Adopt default 1.7 +/- 1.7 c/s.
2a.  MEB side 2 with wrong SBC filter map.
2b.  MEB side 2, correct SBC filter map?
3.  No lamp filter.
4.  2.5mm pinole.

Table 2b (old). Throughput measurements of Xe 1469Å (no bad row replacement)
Entry # MJD -
51200
SBC
filter
texp
[s]
global rate
[counts/s]
SBC/PMT
(rate ratio)
notes
Splat dark
3475 3476 29.12631 F115LP 100 1569 ± 3 0.967 ± 0.003 1  
3477 3476 29.13228 F115LP 100 1571 ± 3 0.961 ± 0.003 1  
3613 3614 31.82258 F115LP 100 1404 ± 3 0.957 ± 0.003 1  
3697 3698 34.71709 F115LP 100 1457 ± 3 0.933 ± 0.003  
3699 3698 34.72633 F115LP 100 1501 ± 3 0.964 ± 0.003  
3831 3832 37.02042 F115LP 100 1166 ± 3 0.939 ± 0.004 2a  
3833 3832 37.02771 F115LP 100 1196 ± 3 0.947 ± 0.003 2a  
4307 4308 39.07912 F115LP 100 992 ± 3 0.938 ± 0.004  
4309 4308 39.08516 F115LP 100 1007 ± 3 0.929 ± 0.004  
4802 4803 42.16699 F115LP 100 767 ± 3 0.954 ± 0.005  
4804 4803 42.17456 F115LP 100 774 ± 3 0.955 ± 0.005  
4860 4861 44.42222 F115LP 100 673 ± 2 0.946 ± 0.005  
4862 4861 44.42916 F115LP 100 674 ± 2 0.950 ± 0.005  
4939 4938 47.22839 F115LP 200 551 ± 2 0.950 ± 0.005 1  
4940 4938 47.23623 F115LP 100 5208 ± 7 0.944 ± 0.002 1  
4941 4938 47.24286 F125LP 300 532 ± 1 0.915 ± 0.005 1  
4942 4938 47.24956 F140LP 300 498 ± 1 0.849 ± 0.005 1  
4943 4947 47.25657 F150LP 300 160 ± 1 0.272 ± 0.008  
4944 4947 47.26288 PR110L 300 495 ± 1 0.833 ± 0.004  
4945 4947 47.26910 PR130L 300 515 ± 1 0.878 ± 0.004  
4946 4947 47.27772 F165LP 300 112 ± 1 0.002 ± 0.009  
5906 5907 51.45113 F115LP 100 4004 ± 6 0.960 ± 0.002 2b  
Notes: see bottom of Table 2a