Instability Theories
Troubleshooting has Confirmed that Optical Bench is not Isolated from Enclosure (XYZ Fittings should Isolate Bench)
- This Anomalous Behavior May Cause/Contribute to All Phenomena through the Following Mechanisms:
- Enclosure Shape Change Transmits Loads to Optical Bench
- TBF/Dolly CTE Effects Load Enclosure which Load Bench
- Optical Bench CTE Expansion Results in Bench Bending
Other Effects Which May be Contributing to Instability:
- Heat Pipe CTE Expansion Imparts Loads to Optical Bench
- Localized CTE Effects in Detector Housings & Other Components Expanding/Contracting CCD Heat Pipes Impart Load to CCDs and Optical Bench