March 30: Verify Reproduce-ability & Correlation w/WFC
Configuration:
- Mounting: ACS in TBF/Diamond Configuration, in SSDIF
- Cooling: Chiller Plate
- Optical Source: WFC: RAS/Cal point source; HRC: Coronagraph
- Starting Temp: Ambient, at Start of Day
Results:
- Little to no motion seen during ACS "warm-up"
- Chillers turned on to 14C causes Interface Plate temperature change with causes motion in both WFC and HRC
- In this test, a 2C delta causes ~ 4 pixels of motion
- HRC moves same direction as Mar 22 data set (parallel to floor)
- WFC moves:
- Perpendicular to floor (up and down) as viewed into aperture
- Fore to Aft at CCD, as viewed looking down on instrument