April 15 PM: Compare Dolly/Square Motion when HPs Disconnected
Configuration:
- ACS: CCD Heat Pipes Disconnected at IF Plate
- Mounting: ACS in Handling Dolly in Square Orientation
- Cooling: Chillers
- Optical Source: HRC: Coronagraph
- Starting Temp: Warm after several hours of Operation
Results:
- Note: data noisy due to operating CCDs at room temperature (TEC off)
- Start warm with chillers on at 14C
- No apparent motion due to turning chillers off
- Turning HRC TEC on at 21.9 causes a large jump due to image quality improvement (not a thermal-mechanical effect)
- Correlation between motion & TEC on/off state identified (due to Housing temp?)
- Turning HRC TEC off at 22.7 causes small ~0.3 pixel motion
- Turning HRC TEC on at 23.0 causes TEC-off motion to reverse