April 16: Compare TBF/Diamond Motion when HPs Disconnected
Configuration:
- ACS: CCD Heat Pipes Disconnected at IF Plate
- Mounting: ACS in TBF in Diamond
- Cooling: Chillers
- Optical Source: HRC: Coronagraph
- Starting Temp: Ambient, at start of day
Results:
- Warm-up effect of ~1.5 pixels seen
- Direction of motion similar to that with heat pipes connected
- Chillers-on seemed to restart warm-up trend which had stalled for ~15 minutes
- Chillers off had no apparent effect
- Motion correlated with TEC on/TEC off repeated