June 4: Measure Stability with Radiator Panel Rebolted
Configuration:
- ACS: Assembled with Panel Bolts Re-Tightened
- Mounting: In Dolly, in Square with Radiator Facing Up & Parallel to Floor
- Cooling: Chillers
- Optical Source: HRC Coronagraph Backlight
- Starting Temp: Ambient at Start of Day
Results:
- Executed Test Sequence Identical to Previous Day for Apples-Apples Comparison
- Resulting Pixel Motion Nearly Identical to Previous Day, Except:
- Warm-up Effect Barely Present
- Possibly because ACS Started 1C Warmer?
- Possibly because Enclosure is More Resistant to Shape Changes with Panel?
- HRC Set Point Change Caused Slope Change
- This Effect Probably Occurred Yesterday but was Masked by Warm-up Effect
- "Rolling" Effect Larger Magnitude
- Rolling Appears Correlated with JIN2PANT Enclosure Panel Temperature
-
- Except for Rolling, Stability is Better with Enclosure Panel Bolts Torqued