June 17: Compression Test to Verify Presence of Y-Fitting "Stiction"
Configuration:
- ACS: Fully Assembled; OFF for this Test to Eliminate Thermal Effects
- Mounting: In TBF/Diamond
- Cooling: Chillers
- Optical Source: None (ACS Off)
- Starting Temp: Ambient throughout Test (ACS Off)
- Add'l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure &
Theodolites to Monitor TBF and ACS Optical Cube Motion
Figures:
- LVDT setup: ( HTML )
- LVDT data: ( PDF )
Results:
- TBF B-Latch Used to Compress Enclosure from 0 to 800lb in 50lb Increments
- Increasing Load Compresses Both Enclosure & Optical Bench by ~25 mil
- ACS Optical Cubes' Motion of ~6 to 18 as Indicates Bench is Bending w/Increasing Load
Conclusion:
- Optical Bench Moves with Enclosure
- Y-Fitting is not Behaving per Design