June 18: Repeat Compression Test for Repeatability
Configuration:
- ACS: Fully Assembled; OFF for this Test to Eliminate Thermal Effects
- Mounting: In TBF/Diamond
- Cooling: Chillers
- Optical Source: None (ACS Off)
- Starting Temp: Ambient throughout Test (ACS Off)
- Add'l Sensors: LVDTs to Monitor Motion of Bench Relative to
Enclosure & Theodolites to Monitor TBF and ACS Optical Cube Motion
Results:
- Decompression Reverses Compression Effect with Little or No Hysteresis
- Compression and Decompression Effects are Repeatable
Conclusion:
- Optical Bench Moves with Enclosure
- Y-Fitting is not Behaving per Design