June 21 AM: Compression Test with ACS On to Correlate Spot Motion
Configuration:
- ACS: Fully Assembled
- Mounting: In TBF/Diamond
- Cooling: Chillers
- Optical Source: HRC Coronagraph Spot with Backlight
- Starting Temp: Ambient at Start of Day
- Add'l Sensors: LVDTs to Monitor Motion of Bench Relative to Enclosure & Theodolites to Monitor TBF and ACS Optical Cube Motion
Results:
- TBF B-Latch Used to Compress/Decompress Enclosure
- Both Enclosure & Bench Move with Changing Load
- Coronagraph Moves ~7 Pixels in Each Axis with Changing Load
- Compression Effect is Repeatable; Decompression is Repeatable
- Compression vs Decompression Motion Shows Hysteresis
Conclusion:
- Coronagraph Motion is Correlated with Enclosure/Bench Motion and is thus Caused (at least in part) by Stuck Y-Fitting