June 21 PM: Compression Test with ACS On to Correlate Spot Motion
Configuration:
- ACS: Fully Assembled
- Mounting: In TBF/Diamond
- Cooling: Chillers
- Optical Source: SBC Coronagraph Spot with Backlight
- Starting Temp: Warm after Morning Operations
- Add'l Sensors: LVDTs to Monitor Motion of Bench Relative to
Enclosure & Theodolites to Monitor TBF and ACS Optical Cube Motion
Results:
- TBF B-Latch Used to Compress/Decompress Enclosure
- Both Enclosure & Bench Move with Changing Load
- Coronagraph Moves with Changing Load
- Compression vs Decompression Motion Show Hysteresis
- SBC Motion is Nearly Identical to HRC Magnitude and Direction
Conclusion:
- Coronagraph Motion is Correlated with Enclosure/Bench Motion and
is thus Caused (at least in part) by Stuck Y-Fitting
- Similarity of SBC & HRC Motion Indicate Effect is Predominately
Motion of the M2 Mirror