Table 2 : HRC Gain - Flat Fields

Cal. Log
Numbers
texp
[s]
THRC
[C]
Filter Plots Gain
[e-/DN]
Comments
5338 & 5339
5340 & 5341
Bias: 5336
1.0
2.0
-72 F625W
F625W
Slope
Histogram
1.139 AMP B,
11 March 1999,
Tungsten-3
5408 & 5409
5410 & 5411
Bias: 5406
1.0
2.0
-84 F625W
F625W
Slope
Histogram
1.122 AMP B,
11 March 1999,
Tungsten-3
5444 & 5445
5446 & 5447
Bias: 5442
1.0
2.0
-80 F625W
F625W
Slope
Histogram
1.122 AMP B,
12 March 1999,
Tungsten-3
5480 & 5481
5482 & 5482
Bias: 5478
1.0
2.0
-82 F625W
F625W
Slope
Histogram
1.128 AMP B,
12 March 1999,
Tungsten-3
5707 & 5708
5709 & 5710
Bias: 5705
1.0
2.0
-92.8 F625W
F625W
Slope
Histogram
1.123 AMP B,
13 March 1999,
Tungsten-3